How to fill out the Measuring Tape Calibration Procedure Pdf online
Filling out the Measuring Tape Calibration Procedure Pdf is a straightforward process that ensures accurate measurement standards. This guide provides clear step-by-step instructions to help users complete the form efficiently, ensuring all necessary details are accurately captured.
Follow the steps to complete the Measuring Tape Calibration Procedure Pdf online.
- Press the ‘Get Form’ button to access the Measuring Tape Calibration Procedure Pdf. This will allow you to open the document in an online editor.
- Begin by reviewing the 'Standards and Equipment' section. Ensure you have the necessary equipment, such as a granite plate and a master gauge block set, along with valid calibration certificates for the standards and equipment.
- Navigate to the 'Calibration Procedure' section. Make sure to use the 'TS-0069 Measuring Tape Calibration Record Sheet' as instructed.
- Clean all relevant measuring surfaces, including the measuring tape, granite surface, and gauge blocks to be utilized.
- In your form, measure the ambient temperature, noting it down. If the recorded temperature is less than 18°C or greater than 24°C, refer to the Division Manager for further guidance.
- Perform a visual inspection of the measuring tape. Slide out one foot of the tape and inspect for damage. Ensure the scale markings and tabbed metal clip are visible, and that the tape moves smoothly along its entire length.
- For inside scale measurement, use gauge blocks to measure and record 5 different lengths. Note that you can join multiple gauge blocks for longer measurements.
- Position the gauge blocks vertically on a granite surface, secured by a height gauge. Make sure to set the metal clip of the measuring tape against the granite surface to ensure accuracy.
- Record the measured lengths and verify they are within the acceptable limits of +/- 0.05 inches (1/16 inch) or 1mm.
- For outside scale measurement, repeat the process as necessary and ensure all measurements align with the established acceptable limits.
- Once all fields are completed and verified, save your changes. You may also download, print, or share the completed form directly from the editor.
Complete the Measuring Tape Calibration Procedure Pdf online to ensure accurate measurements and compliance with standards.
The five points of calibration include defining the purpose, identifying measurement standards, outlining the process, documenting results, and establishing a schedule for regular checks. Each point plays a crucial role in ensuring that the measurements are consistent and reliable over time. A well-prepared 'Measuring Tape Calibration Procedure Pdf' will cover these aspects thoroughly. For assistance in developing these procedures, consider visiting our platform for expert resources.
Industry-leading security and compliance
-
In businnes since 199725+ years providing professional legal documents.
-
Accredited businessGuarantees that a business meets BBB accreditation standards in the US and Canada.
-
Secured by BraintreeValidated Level 1 PCI DSS compliant payment gateway that accepts most major credit and debit card brands from across the globe.